Page 96 - FYSIOCatENG08-2018
P. 96

fysiomed
20110 - PRO-KIN PK 252
The Pro-Kin system offers an excellent proprioceptive stabilometric assessment.
The stability of the mobile platform is managed via a sophisticated electro-hydraulic which allows a fluid and precise movement of the platform.There are 50 instability levels, for a very gradual bipodalic proprioceptive training.The levels from 1 to 10 are very soft and allow also monopodalic exercises.
The user friendly software, equipped with a touch screen, makes managing the data of analysis, assessment and training easy.The main menu of the software facilitates the access to all the principal functional modules of the Pro-Kin. Provided with a number of assessment programs, it enables the operator to understand correctly the proprioceptive condition of the patient. After the assessment, the software automatically suggests a series of rehabilitation tracing, designed to improve the perceptive conditions of each single movement.
The rehabilitation tests and protocols are stored in the patient file program and can be called up anytime for further control and comparing tests.
The Pro-Kin 252 can be blocked monoaxial.The front-behind axis of the platform can be blocked leaving the left-right axis free and vice versa.The monoaxial exercise is particular suitable for every proprioceptive rehabilitation itinerary. Only when the patient is able to manage correctly both the axes individually, it’s possible to pass to a proprioceptive bipodalic multiaxial training.
The Pro-Kin is equipped with 5 load cells and offers the possibility of transformation into a complete stabilometric system. By closing off the pneumatic circuit flow interrupters, the mobile platform accurately stabilizes itself, thus al- lowing a classic type of orthostatic stabilometric assessment (Romberg Test, C.O.P. Oscillation, ...)
Also innovative is the instability control system, whereby thanks to a feed-back reaction device, the mobile platform stability is automatically adjusted in relation to the mass of the subject being assessed and to the chosen instability coefficient.
The TK Trunk Sensor is another innovative feature of the Pro-Kin 252: a biaxial clinometer capable of detecting trunk oscillations in anterior-posterior and left-right directions.Trunk stability during proprioceptive exercises in erect pa- tient position, indicates that the subject has good peripheral control (ankle, knee, hip) and does not require any pos- tural compensation at higher levels.
Assessment modules:
• proprioceptive pre-assessment
• monoaxial proprioceptive assessment • multiaxial proprioceptive assessment • disequilibrium/skill assessment
• time/ dynamic assessment
• equilibrium management assessment
Parameters:
• 15° degrees of inclination from horizontal position to all directions. 0.1° measure reading maximum resolution • electric supply: 230 VAC, 50/60 Hz
• absorbed power: 400 Watts
• medical decouple transformer 230v/24v P.nom. 600 VA
• base: 150 cm x 160 cm
• height: 180-200 cm
• weight: 95 Kg
• measures reading sampling: 20 Hz
• patient’s load capacity: 120 Kg (135 Kg safety)
• certifications: EN 60601- 1 (CEI 62- 5), 93/42/CEE, Class II - Type B a
Standard:
• touch screen display 17” with 1024 x 768 pixel resolution
• control unit: processor, QWERTY keyboard, optical mouse, serial communication port RS232, S.O. windows XP
Multi-language
• adjustable display height
• dynamic/static platform with 55 cm diameter
• support platform and assistance 110 x 110 cm, ground elevation of 21 cm.
• trunk sensor
• wrapping upper supports
• manuals: operator’s manual (paper and digital), software manual (digital), clinic manual (digital), rehabilitative pro-
tocols (ankle, knee, hip)
Not included: colour printer inkjet HP Desk-jet
94