Information on reprocessing of instruments according to resterilizable DIN EN 17664, as well as indications for labeled as disposable products
PRODUCT
Manufacturer's information applies to all rotary instruments, which are delivered by Diamond D&G. These instruments include: diamond drills, crown cutters, tungsten carbide burs, finishing burs, diamond wheels, ceramic ganders and polishers, Information on merchandise like endodontic products are to request from the appropriate vendor. Once items are opened K H not suitable for reuse.
INSTRUCTIONS FOR PREPARING
Please note that in principle, only equipment and product specific validated methods for cleaning, disinfection and sterilization may be applied. The equipment used (ultrasonic bath, RDS and sterilizer) must be regularly checked and maintained. This is the responsibility of the practice owner or the hygiene officer. Cleaning and disinfection of all instruments (semi-critical A • B, A B critical).
AUTOMATED REPROCESSING
Thus, the disinfected instruments do not damage each other, they should be positioned in appropriate instrument blocks. The autoclavable instrument block must be placed properly in the disinfection machine so that the water comes directly on the instruments. Use the right dose according to the manufacturer suggests. Now start the program with thermal disinfection. Make sure that the program is correct.
Upon completion of the program, remove the instruments and dry them completely with the help of compressed air. Control the instruments for cleanliness, surface damages and form damage. Damaged instruments must be discarded directly. If there is visible residual contamination, repeat the process until all the impurities removed. The process until all the impurities removed.
MANUAL REPROCESSING
Provide the instruments in the appropriate column in an ultrasonic bath. Please refer to the manufacturers instructions to concentration and exposure time. Warning: Polisher and ceramic grinding tools are not allowed in an ultrasonic bath. The resulting can be vibrations, loss of materials and damages. This exclusively in but bath (non- alcoholic, for example saniswiss bio sanitizer I) process.
STERILIZATION
Instruments of class semi-critical A • B need not necessarily be sterilized. For more information, see the Robert Koch Institution guidelines.
Packaging
Instruments of class critical A • B must be sealed in suitable disposable sterilization packaging. It is important to ensure that the seal is not under tension and has no wrinkles.
Steam sterilization according to Din EN 13030 class B Sterilize according to the equipment manufacturer.
Ø Sterilization temperature 134 C;
Ø Fractionated pre-vacuum;
Ø Holding time at least 5 minutes (full cycle);
Ø Drying time min 10 minutes;
STORAGE
The storage of all instruments (sterile and non-sterile) must be clean, dust and moisture free. In addition, recontamination must be excluded.
MAINTENANCE
Rotary Instruments require no maintenance.