Page 71 - M1-12 Ops Manual McMurdo 2016
P. 71

P (L1) N (L2)
D) POWER LINE WIRING
and within easy reach of the operator;
- it shall be marked as the disconnecting device for the equipment.
NOTE: a single switch or circuit-breaker can drive more than one instrument.
4
N (L2)
5
P (L1)
POWER LINE 100 V to 240 V A.C (50/60Hz)
or 24 V AC/DC
9) When a neutral line is present, connect it to terminal 4
Fig. 13 POWER LINE WIRING
NOTES:
1) Before connecting the instrument to the power
line, make sure that line voltage corresponds to
the descrtiption on the identification label.
2) To avoid electrical shock, connect power line at
the end of the wiring procedure.
3) For supply connections use No 16 AWG or larger
wires rated for at last 75 °C.
4) Use copper conductors only.
5) Don’t run input wires together with power cables.
6) For 24 V DC the polarity is a do not care condition.
7) The power supply input is NOT fuse protected.
Please, provide it externally.
Power supply 24 V AC/DC 100/240 V AC
Type Current T 500 mA T 125 mA
Voltage 250 V 250 V
When fuse is damaged, it is advisable to verify the power supply circuit, so that it is necessary to send back the instrument to your supplier.
8) The safety requirements for Permanently Connected Equipment say:
- a switch or circuit-breaker shall be included in the building installation;
- It shall be in close proximity to the equipment
1604-1-BC.p65 8
5/16/00, 10:25 AM
8