DEFAULT CONFIGURATION
P ARAMETERS
The configuration parameters can be loaded with predetermined default values. These data are the typical values loaded in the instrument prior to shipment from factory. To load the default values proceed as follows:
tb1 dFLt
a) The internal switch should be open.
b) The upper display will show:
LOAd
1604-A-BC.p65
2
5/16/00, 10:25 AM
COnF
It means that the loading procedure has been initiated.
c) Push the t pushbutton; the display will show the firmware version.
After about 3 seconds the loading procedure is terminated and the instrument reverts to visualization as in point b.
C0nF A. 0 0
SEr 1 ErO ErO SEr2 1 1
d) Maintaining the pressure on the t pushbutton, push the s pushbutton also. The instrument will show:
OFF dFLt
e) Press s pushbutton to select between table 1 (European) or table 2 (American) default set parameters; the display will show:
A. 2
f) Press FUNC pushbutton; the display will show:
PARA. TABLE 1 TABLE 2
SEr SEr P1 P2 P3 P4 P5 P6 P7 P8 P9 P10 P11 P12 P13 P14 P15 P16
3 4
19.20 19.20 7E 7E 3 20 ----. ----. 0 0 400 1000 rEL rEL rEV rEV 1 1 AIr AIr H.A. H.A. OFF OFF 10 10 0 0 H.A H.A. OPrt. OPrt 0 0 0.1 0.1