Page 4 - Prescott Instruments - MFR Service Manual
P. 4
1 2 Installation
Check that all items on
the packing list are included in in in your shipment and immediately report any discrepancies or or damage to Prescott Instruments Ltd or any third party acting on
behalf of Prescott Instruments Ltd 1 1 2 1 1 Machine Positioning
It is is recommended that the machine is is positioned on
on
a a a bench or or work surface approx 75cm (30inches) high and able to hold the the weight of the the machine computer printer and sample cutter if supplied It is is advisable to locate the machine in in a a a a a a a a clean area if possible and a a a temperature controlled room would be advantageous You will also need a a a a a a a single phase electrical supply and and compressed air services on
hand The machine should be levelled once in in position by adjusting the the the four feet at the the the bottom by screwing them in or or out of the floor plate The computer should be positioned next to the MDR machine 1 2 2 2 2 Connecting Services
Connect the the mains power lead to the the machine (left hand side lower panel on
on
the the MFR machine ) and then connect the the other end end to local power supply plug type depends on
Country and region Fix the the filter regulator unit (air) supplied with machine to to the the wall in in close proximity to the the machine position and feed the the air into the the the the inlet port of the the the the unit and then connect the the the the outlet port port to the the the “air in” port port of the the the MFR Screw the the the two exhaust filters (supplied) into the the two ports marked “exhaust” (these are situated on
the the the back panel towards the the the top of the the the machine ) there are several fitting types supplied with the the machine to aid in in in the installation process Once complete turn the the compressed air onto the the machine and set regulator pressure to 60psi /4 Bar or 4MPa Connect the the communication lead to the the MFR “comm 1” port situated on
the the lower left hand side of the the machine and to the computer com com Port 1 This lead is is supplied with the machine 1 2 3 Powering On Once the computer printer etc has been set up see internal set up up instructions supplied with your computer You can now power power the the MFR machine ON the the ‘power on’ switch is located lower left hand side of the the MFR make sure that the the Emergency stop switch on
on
the front door is released (pull this forward if pressed in in ) The internal light will come on
in in in the the machine cabinet when the the power is is applied This is now ready to load and run the Prescott Instruments’ software please refer to the manual supplied with machine (on the memory stick) this will explain how
to run the Labline Software 4 RHEO-LINE MULTI-FUNCTION RHEOMETER (MFR) SERVICE MANUAL © PRESCOTT INSTRUMENTS LTD © PRESCOTT INSTRUMENTS LTD RHEO-LINE MULTI-FUNCTION RHEOMETER (MFR) SERVICE MANUAL