Page 42 - PBL71 OI/AXM630-EN AZTEC 600 SS (WIRO)
P. 42
Aztec 600 ISE ammonia and fluoride
Single-stream ion-selective analyzers
6.5 Current Outputs
There are seven tabs in the Outputs screen, one for each output and an output calibration tab:
6.5.1 Outputs 1 to 6
6 Configuration
Fields
Description
Output Source
The 'Output Source' field can be set to one of the following options: None – no other fields are visible
Stream 1
The following fields are displayed only if 'Output Source' is set to 'Stream 1':
Output Range
The low and high limits for the output range. Both values can be set independently. If the difference between the low and high limits is too small, the output is very noisy.
Output Type
The electrical low and high limits (0 to 22 mA).
For example, if the low and high limits for 'Output Range' are set to 0 and 1000 mg/l respectively, and the electrical low and high limits are set to 4.00 and 20.00 mA, at 0 mg/l the output is 4.00 mA and at 1000 mg/l the output is 20.00 mA.
Out of Sample Ind.
Out of sample indicator. If set to 'Yes', the output goes to the default output value when an out of sample condition occurs for the selected stream source.
The following field is displayed only if 'Out of Sample Ind.' is set to 'Yes':
Default Output
The output value used when an out of sample condition occurs and 'Out of Sample Ind.' is set to 'Yes' (0 to 22 mA).
6.5.2 Output Calibration
Fields
Description
Calibrate Output 1 (to 6)
Enables each output to be calibrated.
38
OI/AXM630–EN Rev. J