Page 62 - PBL71 OI/AXM630-EN AZTEC 600 SS (WIRO)
P. 62

Aztec 600 ISE ammonia and fluoride
Single-stream ion-selective analyzers
12 Diagnostic Information and Icons
12.1 Analyzer Diagnostic Information
The analyzer is programmed to display diagnostic messages to provide information on servicing requirements and any other conditions that develop during operation.
All diagnostic messages displayed on the analyzer are added to the analyzer's audit log.
Table 12.1 shows icon types, diagnostic messages and ON / OFF status for the fixed relays, all of which operate in the fail-safe condition.
12 Diagnostic Information and Icons
  Diagnostic Icon
NAMUR Status
 Failure
 Check function
 Out of specification
 Maintenance required
           Note. The diagnostic icons in the following table conform to NAMUR 107.
    Icon
Diagnostic message
  Stop Relay
Attention Relay
  Calibration Relay
Failure Relay
 Monitor in Service
The analyzer is measuring correctly.
ON
ON
ON
ON
 20 Character User-defined Alarm Tag
User-defined alarm message.
—
—
—
—
 Sample Flow Failure
The analyzer cannot detect any sample flow.
If there is sufficient flow to the analyzer, check that:
1. the float is located in the side-sample pot and is unrestricted.
2. the end cap of the side-sample pot is positioned to allow the magnetic float to be within the area of operation of the reed switch.
3. the float switch lead is connected correctly to the interconnection board.
4. the magnetic reed switch is operating correctly.
ON
OFF
ON
ON
 Monitor Stopped
Analyzer operation has been stopped by the user.
OFF
ON
ON
ON
 Monitor Off
Analyzer operation has not been started.
ON
ON
ON
ON
 Monitor Off
A failure has caused the analyzer to stop operation and switch off all services.
ON
ON
ON
OFF
 Priming
The analyzer is priming the tubing. The priming routine draws in each reagent, sample and standard in turn, filling the tubing and then pumping to waste.
An automatic calibration starts when the priming routine is completed.
OFF
ON
ON
ON
               Table 12.1 Diagnostic Information, Relay Operations and Inhibits (Sheet 1 of 4)
58
OI/AXM630–EN Rev. J