Page 93 - PBL71 OI/AXM630-EN AZTEC 600 SS (WIRO)
P. 93
Aztec 600 ISE ammonia and fluoride
Single-stream ion-selective analyzers
Appendix F β Updating the Software
Appendix F β Updating the Software
9.
A 'Warning' screen appears next. This provides the option to upgrade the existing system software or exit without upgrading:
Normal Operation (Up)
Software Update (Down)
C
B A
Fig. F.1 SD Card Removal
1. Stop analyzer operation by selecting 'Stop Monitor' from the main operator menu.
2. Ensure the analyzer is switched off.
3. Use a large flat-headed screwdriver to release the two
door catches A.
4. Remove the SD card B.
5. Refer to the documentation on the web site (http://www.abb.com) for details of any software updates.
6. Download the software and transfer it to a blank SD card using an appropriate card reader.
7. Set the switch C (above the SD card socket) to the lower position (Software Update).
8. Insert the SD card, close the door and switch on the analyzer. A start-up message appears for a short duration (approximately 5 seconds) followed by a 'File Load Progress' bar then the 'ABB' splash screen:
10. To proceed, press the key.
11.The 'Commissioning V2.0' screen appears with the 'Upgrade Firmware' button selected:
12.To upgrade the existing system software, press the
key to start the process. The software is transferred to the analyzer in sections.
13. Press the key as each section is installed (installation may take a few minutes).
14. Open the door and move the switch to the up position.
15. Close the door and re-tighten the door catches (A in Fig. F.1).
16. Turn power off and then on to restart the analyzer.
OI/AXM630βEN Rev. J