Page 62 - Sterilization Training Manual
P. 62

May 31, 2017
Subject: H2OWin Software / SEC PC IR Link Product: SEC Signature DIR
The H2OWin software, developed by Sensor Electronics Corporation (SEC) is designed to be used with the SEC PC IR Link and a customer supplied personal computer (PC). The software sends and receives information from the computer to SEC Signature DIR EtO/H2O Monitor. The primary use for the H2OWin software is for troubleshooting and calibration of the SEC Signature DIR Monitor. The H2OWin software is not used with the SEC Signature DIR monitor continuously. The software has no real-time data gathering or information storage features.
The SEC PC IR Link is used as a bidirectional communication interface between the SEC Signature DIR and personal computer. There are no calibration requirements for the SEC PC IR Link. If the SEC PC Link is used to power the SEC Signature DIR, the 24 VDC power supply output should be measured annually. The voltage can be measured on Terminals 2 (+) and 3 (-).
Alan H. Petersen Jr.
Sensor Electronics Corporation
Sensor Electronics Corporation 12730 Creek View Avenue Savage, Minnesota 55378 U.S.A. (800) 285-3651  (952) 938-9486  FAX: (952) 938-9617 www.sensorelectronic.com