Page 11 - ZVPlasty - Hensler Technique Guide 2021
P. 11

ST-007 rev-4


             Step 4:  Creation of Void
                                                           Step 4
             Insert the curved cannula through the access cannula at
             a desired depth using markings on curved cannula and
             using imaging as a visual aid. Once desired depth is
             met, rotate the curved cannula back and forth
             (clockwise and counter-clockwise) several times to
             ensure a void is created.





























             Step 5:  Curved Stylet Removal
                                                           Step 5
             After void is created. Remove curved stylet handle by
             rotating counter-clockwise and simultaneously pulling
             the curved stylet handle outward.
















                                              Page 11 of 19
   6   7   8   9   10   11   12   13   14   15   16