Page 4 - ZVPlasty - Hensler Technique Guide 2021
P. 4
ST-007 rev-4
• Table 2 gives the maximum recommended inflation
pressure and volume values in constrained and
unconstrained conditions for 10G, 11G, and 13G. UNCONSTRAINED TEST CONDITION
Also included are the pressure and volume values at Max. Recommended Max. Recommended
burst in constrained and unconstrained conditions for Catalog No. Inflation Pressure Inflation Volume
guidance. VCF-1000-10 6.5 ATM 4 ml
Inflated Balloon VCF-1000-15 6.5 ATM 4 ml
Dimensions 10G
VCF-1000-20 6.5 ATM 6 ml
Catalog No.: VCF- Inflated Dimensions
1000-10
Size (Deflated
Balloon Length): 15 mm Volume Diameter Length CONSTRAINED TEST CONDITION (Please note that the Inflation volume
Max. Inflation 4 ml 2 ml 10 mm 12 mm required to achieve a desired pressure in a constrained condition will vary during
Volume: product use due to local variation in bone structure )
4 ml 16 mm 17 mm Max. Recommended Max. Recommended
Catalog No.
VCF- Inflation Pressure Inflation Volume
Catalog No.: Inflated Dimensions
1000-15 VCF-1000-10 27 ATM 4 ml
Size (Deflated
Balloon Length): 15 mm Volume Diameter Length VCF-1000-15 27 ATM 4 ml
Max Inflation
Volume: 4 ml 2 ml 9 mm 15 mm VCF-1000-20 27 ATM 6 ml
4 ml 14 mm 21 mm
• Inflating the vertebral balloon beyond the maximum recommended inflation
VCF- volume may cause the balloon to rupture before reaching the maximum
Catalog No.: Inflated Dimensions
1000-20 inflation pressure and
Size (Deflated 18 mm Volume Diameter Length • Inflating the vertebral balloon beyond the maximum recommended inflation
Balloon Length): pressure may cause the balloon to rupture before reaching the maximum
Max Inflation inflation volume.
Volume: 6 ml 2 ml 8 mm 18 mm
4 ml 12 mm 24 mm
6 ml 16 mm 29 mm
UNCONSTRAINED TEST CONDITION
Inflated Balloon Max. Recommended Max. Recommended
Dimensions 11G Catalog No. Inflation Pressure Inflation Volume
Catalog No.: VCF- Inflated Dimensions VCF-1100-10 6.5 ATM 4 ml
1100-10
Size (Deflated VCF-1100-15 6.5 ATM 4 ml
Balloon Length): 15 mm Volume Diameter Length
Max. Inflation 4 ml 2 ml 10 mm 12 mm VCF-1100-20 6.5 ATM 6 ml
Volume:
4 ml 16 mm 17 mm
CONSTRAINED TEST CONDITION (Please note that the Inflation volume
VCF-
Catalog No.: Inflated Dimensions required to achieve a desired pressure in a constrained condition will vary during
1100-15 product use due to local variation in bone structure )
Size (Deflated 15 mm Volume Diameter Length
Balloon Length): Catalog No. Max. Recommended Max. Recommended
Max Inflation Inflation Pressure Inflation Volume
Volume: 4 ml 2 ml 9 mm 15 mm VCF-1100-10 27 ATM 4 ml
4 ml 14 mm 21 mm
VCF-1100-15 27 ATM 4 ml
VCF-
Catalog No.: Inflated Dimensions
1100-20 VCF-1100-20 27 ATM 6 ml
Size (Deflated 18 mm Volume Diameter Length
Balloon Length): • Inflating the vertebral balloon beyond the maximum recommended inflation
Max Inflation 6 ml 2 ml 8 mm 18 mm volume may cause the balloon to rupture before reaching the maximum
Volume:
inflation pressure and
4 ml 12 mm 24 mm • Inflating the vertebral balloon beyond the maximum recommended inflation
pressure may cause the balloon to rupture before reaching the maximum
6 ml 16 mm 29 mm inflation volume.
Page 4 of 19