Page 6 - Safeview - Carpal Tunnel Release - HSP 2019
P. 6

Retrograde Approach

   • Draw Kaplan's cardinal line from the ulnar aspect of the thumb toward the hook
    or the   Draw a second longitudinal line proximally lrom the third web space.
   • Create   1cm longitudlnal Incision centered at the Junction of these anatomic landmarks.

   • Identify the palmer aponeurosls and enlarge the aponeurotlc opening, laking care
    lo avoid Injury to the vascular arch al the distal edge of the Incision.
   • Divide the distal portion of the transverse carpal ligament.
   • Rasp the undersurface of the transverse carpal ligament lo clear away synovlal tissue.
   • Dilate the ca,rpal tunnel space with sequential dilators.

   • Insert the Safe View ™    Maintain posterior pressure on the hub of the cannula
    to preserve its position beneath the ligament.
   • Insert a 4mm 30 standard arthroscope, and visualize the ligament.
                 °
   • Insert the forward cutting knife through the hub of the   Retract the skin distally, and
    engage the distal cut edge of Iha   Divide under direct vlsuallzotlon.
   • Alm for the wrist crease just ulnar to the palmarls longus tendon.
   • Skin closure is achieved in the usual fashion. Apply a soft dressing.
   SafeView              ®

   Components


                                                  • Forward cutting blade




        • SafeView  cannula - 360° transparent visualization
                  ®
        • Built-in guided track system for
          precise cutting and safety



















            • In-situ Rasp


                                                                                        • Synovial elevator



                                                                                • Sequential dilators



                                Part Number                       Description
                                 I-ISP-1601SV    SafeVlew"·' Endoscopic Carpal Tunnel Release Kit
   1   2   3   4   5   6   7   8