Page 9 - Hensler-Zavation Normandy EPS - 2019
P. 9

Normandy VBR System Surgical Technique Guide ST-017, rev 1
             Step 8 – Implant Insertion and

             Expansion


             Insert the implant into the prepared space and
             center to the endplates of the adjacent
             vertebral bodies.




             Rotate the large expander knob of the inserter
             instrument to expand the implant to the
             desired amount of distraction.  Height is
             shown on the post of the expandable implant.





             Remove the inserter instrument by
             unthreading the attachment shaft from the
             implant by rotating the shaft
             counterclockwise.










             Step 9 – Implant Positioning


             Implant position can be adjusted by use of the
             tamp.  To prevent damage to the endplates,
             reduce the height of the implant until the teeth
             disengage the endplates, reposition the
             implant and then re-expand the cage.



















                  DCR 614                                 CONFIDENTIAL                              Page 13 of 20