Page 88 - Catalogue - cuvettes and solid-state calibration standards
P. 88
UV/VIS photometric accuracy, wavelength accuracy and stray light performance
Part Number Description Price
FF-CALKIT-4 Calibration kit for UV/VIS photometric accuracy, wavelength accuracy and stray light pereformance 3,525.26
Capabilities:
Checking photometric accuracy in the UV/VIS range
Checking wavelength accuracy in the UV/VIS range
Checking stray light performance in the UV/VIS range
Delivery includes:
For photometric accuracy:
FF-HF-0.2 calibration standard 0.2AU (60% transmission)
FF-HF-0.5 calibration standard 0.5AU (30% transmission)
FF-HF-1.0 calibration standard 1.0AU (10% transmission)
FF-FUV-0.5 calibration standard 0.5AU (30% transmission)
For wavelength accuracy:
FF-WAV-1-UV/VIS Holmium Oxide glass calibration standard
FF-WAV-7-UV/VIS Didymium glass calibration standard
For stray light performance:
FF-FUV-0.5 calibration standard 0.5AU (30% transmission)
Other items:
FF-A30 holder
Case
Specifications:
Tolerance of the photometric accuracy filters: +/- 0.0022AU
Tolerance of the wavelength accuracy filters: +/- 0.2nm
External dimensions of the individual filters: 12.5 x 12.5 x 45.0mm
NIST-traceable
80