BASIC TROUBLE SHOOTING GUIDELINES
HOWEVER IF YOU ARE NOT ABLE TO RESOLVE THE PROBLEM
PLEASE CALL 877-290-8636 OPTION 2
PROBLEM POSSIBLE CAUSE SOLUTION
Sore or dry eyes
Air blowing into eyes Slightly tighten top straps to bring mask into
bridge of nose
Mask leaks Mask/headgear not Re-apply mask and adjust headgear. Lie in
positioned properly bed with mask on and pressure on. Keeping
eyes closed, release top Velcro straps, allow
mask to leak. Pull straps just until mask leak
stops and set straps. Repeat with lower straps
Cushions/pillows If you have not replaced your
worn cushions/pillows in the last 3 months, call to
order supplies. (866) 721-2340
Mask size not Contact Pulmonary Solutions to be re-fitted
appropriate for mask
PROBLEM POSSIBLE CAUSE SOLUTION
Air from unit is Heated humidifier Turn down humidifier setting
warmer than normal may be turned up
too high
Filter obstructed by Replace and/or clean filter
debris
Filter or vent may be Move unit away from any materials that may
blocked by obstruct filters or vent to allow good air flow
bedclothes or
drapes
Room temperature Turn down the thermostat at night
is too warm
Unit seems loud Unfamiliar night time Consider soft music, white noise or fan
noise
Unit seems louder
than usual Air leaking from Turn off machine and make sure humidifier
humidifier chamber is placed correctly
connection
Hose is not Check and/or replace hose-, recommended
connected properly replacement is every 3-6 months. Keep
or has hole in it protected under bedding