Page 33 - 25269_ 1
P. 33
#
#25269-EYAL - 25269-EYAL.1.1 | 2 - A | 18-05-15 | 13:52:37 | SR:-- | Black 25269-EYAL - 25269-EYAL.1.1 | 2 - A | 18-05-15 | 13:52:37 | SR:-- | Yellow 25269-EYAL - 25269-EYAL.1.1 | 2 - A | 18-05-15 | 13:52:37 | SR:-- | Magenta 25269-EYAL - 25269-EYAL.1.1 | 2 - A | 18-05-15 | 13:52:37 | SR:-- | Cyan
Instructions d’utilisation (étape par étape) Warranty
2. Retirez le capot du compartiment puis l’ancienne pile.
# Problem Possible cause Solution
3 Short circuit due to Dry the access cavity with
excess liquid a cotton pellet / air-blower.
(irrigation solution, In case of excess bleeding
saliva, blood) in the wait until it has stops.
pulp chamber.
A direct contact of
Display reaction is erratic:
“0.0’’ or “OVER’’ appear on the file with the For isolation:
7 the screen before the gingiva or gingival • adequate preparation
apical constriction is proliferations, e.g. a filling of access cavity;
reached. fractured metal • use a rubber dam.
crown.
A direct contact of
Fig. 6
the file with metal Isolate the file by inserting
restorations (crown, it into a small polyvinyl tube
3. Insérez une pile neuve dans le compartiment en respectant les parapulpal post, before use.
polarités. amalgam filling).
4. Refermez le compartiment de la pile, fixez le capot à l’aide de la vis et
replacez le cache en silicone sur la vis.
La pile neuve doit être chargée pendant 24 heures avant la première utili- 11 Warranty
sation de l'appareil.
Propex Pixi™ is warranted for 12 months from the date of purchase.
The accessories (cables, etc.) are warranted for 6 months from the date
of purchase.
The warranty is valid for normal usage conditions. Any modification or
accidental damage will void the warranty.
12 Disposal of the Product
Recycling: PLEASE DO NOT THROW AWAY! This product and all its
components must be recycled through your supplier.
50/222 B ML PXEU DFU PRT / Rev.12 / 02-2018 (Old ZF 1902128.X) B ML PXEU DFU PRT / Rev.12 / 02/2018 (Old ZF 1902128.X) 35/222
Y-MK1-853
4