Page 21 - 27771-LASTPAGES
P. 21

14




         Appendix
 Specifications for enclosure port immunity to RF wireless communications
 equipment
                  ELECTROMAGNETIC COMPATIBILITY (EMC)
 The Product is intended for use in an electromagnetic environment in which radiated
 RF disturbances are controlled. The customer or the user of the Product can help   The Product is intended for use in professional healthcare facility or
 prevent electromagnetic interference by maintaining radiation levels of RF wireless   home healthcare electromagnetic environment specified in this
 communications equipment (emitters) within the compliance limits specified below.  Appendix. The customer or the user of the Product should assure that
                  it is used in such an environment.
 Recommended radiation levels of RF wireless Communications Equipment
 Minimum
 IEC 60601-1-2    Changes or modifications to this Product not expressly approved by the
 Frequency band  Compliance level  separation
 Test level       manufacturer may result in increased emissions or decreased immunity
 distance
                  performance of the Product and could cause EMC issues with this or
 380 – 390 MHz  27 V/m  27 V/m  0.3 m
                  other equipment. This Product is designed and tested to comply with
 430 – 470 MHz  28 V/m  28 V/m  0.3 m  applicable regulations regarding EMC and shall be installed and put into
 704 – 787 MHz  9 V/m  9 V/m  0.3 m  service according to the EMC information stated as follows.
 800 – 960 MHz  28 V/m  28 V/m  0.3 m
 1 700 – 1 990 MHz  28 V/m  28 V/m  0.3 m  Warning
 2 400 – 2 570 MHz  28 V/m  28 V/m  0.3 m  Use of portable phones or other radio frequency (RF) emitting
 5 100 – 5 800 MHz  9 V/m  9 V/m  0.3 m  equipment near the Product may cause unexpected or adverse
                  operation.
                  Warning
                  The Product must not be used adjacent to, or stacked with, other
                  equipment. If adjacent or stacked use is necessary, the Product must
                  be tested to verify normal operation in the configuration in which it is
                  being used.
























 220/222  BMLPXEMDFUPRT / Rev.12 / 10-2018  #  BMLPXEMDFUPRT / Rev.12 / 10-2018  27771-EYAL - 27771-EYAL.1 | 7 - A | 19-01-08 | 13:43:44 | SR:-- | Yellow   27771-EYAL - 27771-EYAL.1 | 7 - A | 19-01-08 | 13:43:43 | SR:-- | Magenta   27771-EYAL - 27771-EYAL.1 | 7 - A | 19-01-08 | 13:43:43 | SR:-- | Cyan   #27771-EYAL - 27771-EYAL.1





















 13




























 13







 A1030 000 000 00
   16   17   18   19   20   21   22   23   24   25   26