       8       Technical Characteristics   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27  6.5.3  Over-Instrumentation

                       ®
       9       ProPex  II Apex Locator Error Code  . . . . . . . . . . . . . . . . . . . . . . 29        A red segment and a warning signal indicate that the file has passed the
                                                                                                         Apex. The cursor on the tooth icon shows “OVER”.
       10      Troubleshooting  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

       11      Warranty   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

       12      Disposal of the Product  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

       13      Identification of Symbols  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32












                                                                                                 6.5.4   Completion of the Measurements
                                                                                                         Before unplugging the Measurement cable from the device receptacle,
                                                                                                         disconnect the lip clip and the connection hook from the patient.
                                                                                                         Move the file stopper to the selected reference point on the tooth.
                                                                                                         Gently remove the file from the canal and measure the apical length
                                                                                                         between the stopper and the file tip.
                                                                                                         As a safety precaution in order to avoid over-instrumentation, it is
                                                                                                         recommended to proceed as follows:
                                                                                                                                                                   ®
                                                                                                         - stop file progression in the canal at the point where the ProPex  II
                                                                                                         indicates ‘APEX’;
                                                                                                         - place the file onto an endodontic ruler and measure the apical
                                                                                                         length;
                                                                                                         - subtract a minimum of 0.5 mm from the measured file length.












       4/172                             B EU PIIL DFU PRT / Rev.07 / 02-2018 (Old ZF 190292.X)  B EU PIIL DFU PRT / Rev.07 / 02-2018 (Old ZF 190292.X)            17/172





















































































                                                                                                                                                                                # 25270-EYAL - 25270_pnim_eyal.1.1 | 1 - B | 18-05-15 | 13:28:55 | SR:-- | Magenta  25270-EYAL - 25270_pnim_eyal.1.1 | 1 - B | 18-05-15 | 13:28:55 | SR:-- |