Page 219 - 27771
P. 219

Compliant Cables and Accessories  Guidance and Manufacturer's Declaration – Electromagnetic Immunity

           The Product is intended for use in the electromagnetic environment specified below.
           The customer or the user of the Product should assure that it is used in such an
 Warning   environment.
 Use of non-original cables or accessories may result in increased
 emissions or decreased immunity performance of the Product.   IEC  Compliance   Electromagnetic environment -
           Immunity test   60601-1-2      level              guidance
 The table below lists cables and accessories for which the   Test level
 manufacturer claims EMC compliance:               Field strengths from fixed RF
                                                   transmitters, as determined by an
                                                                         a
                                                   electromagnetic site survey ,
                                                   should be less than the
 Description  Details                              compliance level in each
                                                                b
 Measurement Cable  Original only                  frequency range .
                                                   Interference may occur in the
 Accessories:
                                                   vicinity of equipment marked with
 Lip Clip  Original only                           the following symbol:
 Connection Hook  Original only
 Original Switching charger only:
 Charger  Input: 100-240 V AC ~ 50-60 Hz  Note
 Output: 6V DC ± 5%, 1000 mA  At 80 MHz and 800 MHz, the higher frequency range
                         applies.
                         Note
                         These guidelines may not apply in all situations.
                         Electromagnetic propagation is affected by absorption and
                         reflection from structures objects and people.
           a
           Field strengths from fixed transmitters, such as base stations for radio (cellular/
           cordless) telephones and land mobile radios, amateur radio, AM and FM radio
           broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
           assess the electromagnetic environment due to fixed RF transmitters, an
           electromagnetic site survey should be considered. If the measured field strength in the
           location in which the Product is used exceeds the applicable RF compliance level
           above, the Product should be observed to verify normal operation. If abnormal
           performance is observed, additional measures may be necessary, such as re-orienting
           or relocating the Product.

           b
           Over the frequency range 150 kHz to 80 MHz, field strengths should be less than
           3 V/m.



 214/222  BMLPXEMDFUPRT / Rev.12 / 10-2018  BMLPXEMDFUPRT / Rev.12 / 10-2018  219/222





 #27771-EYAL - 27771-EYAL.1 | 7 - B | 19-01-08 | 11:18:17 | SR:-- | Black 27771-EYAL - 27771-EYAL.1 | 7 - B | 19-01-08 | 11:18:17 | SR:-- | Yellow 27771-EYAL - 27771-EYAL.1 | 7 - B | 19-01-08 | 11:18:17 | SR:-- | Magenta 27771-EYAL - 27771-EYAL.1 | 7 - B | 19-01-08 | 11:18:17 | SR:-- | Cyan   #
   214   215   216   217   218   219   220   221   222   223   224