The selection, use, maintenance and quality control of laboratory equipment and supplies   87


                                                                     Figure 2.50  Working principle
                                                                     of an oil immersion objective.
                                                                     Photo: Idzi Potters, ITG, Antwerp,
                                                                     Belgium.

































                2.5  Specialized technology              range of diagnostic tests-kits for veterinary dis-
                                                         ease diagnosis at the district and regional level.
                ELISa                                      The ELISA plate reader is a rather expensive
                                                         but robust piece of equipment, which provides
                The technology associated with the ELISA   the opportunity to expand the technology sig-
                (enzyme linked immunosorbent assay) has   nificantly. It is basically a spectrophotometer
                made diagnostic serology and antigen capture   that is able to rapidly read and record colour or
                tests much more accessible for smaller veteri-  turbidity in 96 wells in a microtiter plate (Figure
                nary laboratories. However, to perform a lot of   2.51 and Figure 2.52). The cheaper models of
                tests it is necessary to use a semi-automated   plate readers use filters to select a wavelength
                ELISA system. This may include a multiplate   (each wavelength needs a specific filter), the
                system, a photoelectric reader, an automated   more expensive readers use a monochromator
                washer, shaker and specialist incubators. Many   (can select a wavelength from 200 to 999 nm)
                ELISA test-kits allow manual processing, but   (Figure 2.53).
                for multiple sampling, a reliable plate reader   It is worth noting that dual-wavelength
                which can be set to the wavelength required for   measurements can compensate for optical
                the test, is needed. The widespread availability   interference. Dual wavelength is used in many
                of enzyme-linked reagents and suitable colour-  microplate-based applications to reduce optical
                linked enzyme systems has resulted in a wide   interference caused by scratches, fingerprints or