Page 251 - Vessel Sanitation Program 2018 Operations Manual
P. 251

VSP 2018 Operations Manual




                                     The UV DISINFECTION system must be maintained and operated in
                                     accordance with the manufacturer’s recommendation. At least one spare
                                     UV lamp must be available.

                          13.8.3.7  System Shutdown
                                     An automatic shutdown must be maintained whereby any failure in
                                     maintaining the required free residual HALOGEN level, pH level, or UV
                                     lamp intensity must cause the water to completely divert from the BABY-
                                     ONLY WATER FACILITY and instead loop back to the compensation
                                     tank. Additionally, this system must be equipped with an audible alarm
                                     that sounds in a continuously manned space, such as the bridge or engine
                                     control room.

                          13.8.3.8  Shutdown and Alarm Testing
                                     The emergency shutdown and alarms systems must be tested monthly.
                                     Testing procedures and results must be recorded.

                       13.8.4  System Cleaning and Disinfection

                          13.8.4.1  Daily Cleaning of Spray Pad Surface
                                     Every 24 hours, the SPRAY PAD surface and any associated features
                                     must be cleaned with an appropriate cleaner. The surface must be rinsed
                                     and disinfected at 50-ppm free residual HALOGEN for 1 minute, or the
                                     equivalent CT VALUE. Ensure that the liquid waste from this process is
                                     not directed to the compensation tank.


                                     At least every 72 hours, the facility must be shut down and these
                                     procedures must be followed:
                                         1. The entire volume of water within the system must be discharged.
                                            This includes the BABY-ONLY WATER FACILITY,
                                            compensation tank, filter housing, and all associated piping.


                                         2. The BABY-ONLY WATER FACILITY, compensation tank, and
                                            filter housing (cartridge filter) must be cleaned with an appropriate
                                            cleanser, rinsed, and disinfected (chlorine or bromine).
                                            DISINFECTION must be accomplished with a solution of at least
                                            50 MG/L (ppm) free residual HALOGEN for 1 minute, or the
                                            equivalent CT VALUE.

                       13.8.5  Monitoring and Record Keeping
                               An automated analyzer-chart recorder capable of recording free residual
                               HALOGEN levels in MG/L (ppm) and pH levels must be installed.

                               The system must be checked for calibration before opening the facility for use,
                               and then every 3 hours thereafter with a test kit accurate to within 0.2 MG/L



                                         Annex 13.8: Baby-Only Water Facility 198
   246   247   248   249   250   251   252   253   254   255   256