Instructions for Cleaning, Sterilization, Inspection and
®
Maintenance of SpineVision medical devices
English version
Application date : 2015-12-09 Reference document: CI_01EN
II. Equipments required
1. Summary
Personal protective equipment as recommended by the cleaning agent supplier
Cleaning bath or vessel large enough to allow complete immersion of the instruments
Non-shedding wipe
1
Tap water
Freshly prepared purified water/ highly purified water
Freshly prepared cleaning solution using a cleaning agent: proteolytic enzyme or aldehyde-free pre-disinfectant
solution (i.e. : STERIS® PROLYSTICA)
Syringe
Soft bristle brushes and bottle brushes for cannulations lumens and other blind areas (ensure that the brushes
has the appropriate shape and size) - Never use metal brushes or steel wool
Absorbent paper
Washer-disinfector with approved efficiency (e.g. CE mark or FDA approval according to ISO 15883), properly
installed, qualified and regularly subjected to maintenance and testing
Steam autoclave
Lubricant (i.e. : Neodisher® IP Spray from Dr. Weigert laboratories)
1
Note : Avoid use of hard water. Use water whose temperature is 22°C to 43°C (73°F to 110°F). Hot water would
cause the binding of proteins and other residues that could affect the efficacy of the cleaning process.
Page 7 of14
This confidential document is the property of SpineVision Do not duplicate without authorization from SpineVision.
Computer filename: CI_01EN.docx