4 Getting Started
4.1 Overview
The following procedure describes how to start up and configure the analyzer prior to operation.
1. Ensure the analyzer is installed and electrically connected correctly – see Section 3, page 10.
2. Ensure the correct reagents / calibration solutions are connected to the analyzer – see Section 3.6, page 17.
3. Ensure sample is connected correctly to the analyzer and that the flow rate, temperature, pressure and particulate size are within the specified limits – see Section 14, page 70.
4. Switch on power to the analyzer.
After an initial power-up period, the main operator screen is displayed.
5. Press the key and use the and keys to select 'Common Configuration' to configure the analyzer:
4 Getting Started
 6.
Press the key and use the and keys to select 'Measurement' to set up the analyzer's measurement parameters:
– Setup – see Section 6.2.1, page 34
– Streams – see Section 6.2.2, page 34
      – – – – – –
Setup – see Section 6.1.1, page 28
Screen – see Section 6.1.2, page 28
Time – see Section 6.1.3, page 29
Security – see Section 6.1.4, page 30
User – see Section 6.1.5, page 33
Operator Messages – see Section 6.1.6, page 33
7.
Press the
'Exit' to exit configuration. A prompt is displayed asking if the current configuration is to be saved:
Press the key to save the configuration to the analyzer’s internal memory.
Press the key and use the and keys to select 'Operate' followed by 'Flush Monitor' and press the key.
Allow the analyzer to flush for a minimum of 1 hour to enable the measurement probe to stabilize.
 key and use the
and keys to select
    8. 9.