Aztec 600 ISE ammonia and fluoride
Single-stream ion-selective analyzers 12 Diagnostic Information and Icons
 Icon
Diagnostic message
 Stop Relay
  Attention Relay
 Calibration Relay
Failure Relay
 No Valid Calibration
There is no valid calibration stored within the analyzer memory.  Calibrate the analyzer.
OFF
ON
ON
OFF
 Non-Volatile Memory Error
There is a problem with either the display electronics or the main board memory.
 Power the analyzer down, wait for 10 seconds and power-up again.
If the error persists, contact the local ABB representative.
ON
ON
ON
OFF
 Probe Failure
The probe is disconnected or the probe has failed.
1. Check the probe connections.
2. Replace the probe.
Restart the analyzer in either Measurement Mode or Calibrate Mode to clear the message.
OFF
ON
ON
OFF
 Reagent Empty
The reagent bottle is empty.
Refill or replace the reagent bottle.
OFF
ON
ON
OFF
             Table 12.1 Diagnostic Information, Relay Operations and Inhibits (Sheet 4 of 4)