Aztec 600 ISE ammonia and fluoride
Single-stream ion-selective analyzers
12.2 Alarm Acknowledge
To acknowledge a particular alarm, use the
12 Diagnostic Information and Icons
Note. Active unacknowledged alarms are identified by a flashing red alarm event icon. Active acknowledged alarms are identified by a continuous red alarm event icon.
To acknowledge all active alarms simultaneously, select 'All' and press the key.
Fig. 12.1 Alarm Example
and
keys to
12.3 Audit Log and Alarm Event Log
The 'Audit Log' and 'Alarm Event Log' are used to display a list of events and alarms identified by icon, sequence number, date and time.
The 'Audit Log' provides an historical log of system activity and the 'Alarm Event Log' provides an historical log of all alarm events in the sequence they occurred.
When the number of entries in each log has reached 500, the oldest data is overwritten by the newest data. Entries are renumbered so that the number of the oldest entry is always 00.
Both logs are accessible from the chart view and bar view – see Section 2.1, page 8 for details of how to navigate to these logs.
12.3.1 Audit Log – Icons
highlight it in the menu and press the
key.
Note. Use the and keys to scroll through log data.
Icon
Event
Power failed
Power restored
Configuration changed
File created
File deleted
Archive media inserted
Archive media removed
Archive media off-line
Archive media on-line
Archive media full
System error / reset archiving
Date / time or daylight saving start / end changed
Security change
FTP logon
Information
Failure – see Section 12.1, page 58
Maintenance Required – see Section 12.1, page 58
Out of Specification – see Section 12.1, page 58
Check Function – see Section 12.1, page 58