Page 66 - PBL71 OI/AXM630-EN AZTEC 600 SS (WIRO)
P. 66

Aztec 600 ISE ammonia and fluoride
Single-stream ion-selective analyzers
12.2 Alarm Acknowledge
To acknowledge a particular alarm, use the
12 Diagnostic Information and Icons
    Note. Active unacknowledged alarms are identified by a flashing red alarm event icon. Active acknowledged alarms are identified by a continuous red alarm event icon.
 To acknowledge all active alarms simultaneously, select 'All' and press the key.
Fig. 12.1 Alarm Example
and
keys to
12.3 Audit Log and Alarm Event Log
The 'Audit Log' and 'Alarm Event Log' are used to display a list of events and alarms identified by icon, sequence number, date and time.
The 'Audit Log' provides an historical log of system activity and the 'Alarm Event Log' provides an historical log of all alarm events in the sequence they occurred.
When the number of entries in each log has reached 500, the oldest data is overwritten by the newest data. Entries are renumbered so that the number of the oldest entry is always 00.
Both logs are accessible from the chart view and bar view – see Section 2.1, page 8 for details of how to navigate to these logs.
12.3.1 Audit Log – Icons
  highlight it in the menu and press the
key.
          Note. Use the and keys to scroll through log data.
    Icon
Event
Power failed
  Power restored
  Configuration changed
  File created
  File deleted
  Archive media inserted
    Archive media removed
Archive media off-line
  Archive media on-line
  Archive media full
 System error / reset archiving
  Date / time or daylight saving start / end changed
    Security change
 FTP logon
  Information
 Failure – see Section 12.1, page 58
  Maintenance Required – see Section 12.1, page 58
  Out of Specification – see Section 12.1, page 58
  Check Function – see Section 12.1, page 58
                         62
OI/AXM630–EN Rev. J