Page 116 - PBL614 IM/SM500F-EN SM500F ( WIRO)
P. 116
SM500F
Field mountable paperless recorder 7 Configuration
Select the frequency with which F0 values are added to the Totalizer log.
Select a 'Log update source' signal (see Appendix A, page 146) – an edge-triggered signal to trigger the addition of the current F0 values to the Totalizer log on a rising edge.
Log update time
Log updated every...
5, 10, 15, 30, 60, 90 seconds
3 minutes
0, 3, 6, 9... etc. minutes past the hour
5 minutes
0, 5, 10, 15... etc. minutes past the hour
10 minutes
0, 10, 20, 30,...etc. minutes past the hour
15 minutes
0, 15, 30, 45 minutes past the hour
20 minutes
0, 20, 40 minutes past the hour
30 minutes
0, 30 minutes past the hour
60 minutes
On the hour
2 hours
Midnight, 2am, 4am, etc.
3 hours
Midnight, 3am, 6am, etc.
4 hours
Midnight, 4am, 8am, etc.
8 hours
Midnight, 8am, 4pm, etc.
12 hours
Midnight, 12am
24 hours
Midnight
Set the required F0 cut off value.
The cut off value is the minimum temperature value (T) used in the F0 value calculation. Temperatures read below this value are ignored.
114
IM/SM500F Rev. AA