Page 32 - PBL614 IM/SM500F-EN SM500F ( WIRO)
P. 32
SM500F
Field mountable paperless recorder 4 Operation
Note.
While in Historical Review mode:
Recording of new data continues unless stopped from the Logging Menu – see Section 5.1, page 47.
The current value is replaced by '– – –' when the historical data is invalid (e.g. when recording was stopped).
Where the trace at the cursor position represents more than one sample, the indicators flash between the maximum and minimum values of those samples.
Menu options remain active – allowing the screen interval to be changed, different scales and channels to be selected, etc.
Operator messages generated are added to the alarm event log at the present time, not the time indicated by the cursor.
All data stored in the recorder's internal memory can be viewed.
The display can be scrolled back to the start of the oldest data.
Archiving to removable media is suspended but all data recorded in the internal memory buffer during this time is archived automatically on exiting Historical Review mode.
Select to move to data stored in the recorder's onboard memory that was recorded at a specific date and time.
Note.
Oldest data in internal memory
'Goto' target date 'Goto' target time
Returns to Menu
Moves cursor between 'goto' target date and time boxes
Newest data in internal memory
Selected 'goto' target date/time is within daylight saving period
Goes to the Chart view at the selected 'goto' target date and time
If daylight saving is enabled (see Section 7.6.3, page 76) and the selected 'Goto' target date/time is within the daylight saving period, 'Daylight Saving' is displayed on the dialog box.
Once internal memory becomes full, the oldest data is overwritten by the newest data. If historical review has been selected for some time, the oldest data present may no longer be available.
The recorder exits Historical Review mode automatically after 15 minutes if no key is pressed.
30
IM/SM500F Rev. AA