Page 74 - PBL614 IM/SM500F-EN SM500F ( WIRO)
P. 74

SM500F
Field mountable paperless recorder 7 Configuration
Referring to Fig. 7.7, page 73, exit configuration mode as follows:
a Open the menu, select 'Exit' and press .
b To begin using a configuration immediately, select 'Save as Current Configuration'.
       Note.
 When saving the current configuration to internal storage, it is saved twice, once with the filename 'SM500F.cfg' and again with the filename '<time><date><instrument tag>.cfg'
 When saving the current configuration to external storage, it is saved with the filename '<time><date><>instrument tag>.cfg'. It is also saved automatically to internal storage with the filename 'SM500F.cfg'.
 When 'Save Configuration' is selected, the configuration file is saved with the filename '<time><date><>instrument tag>.cfg' to either selected location, internal or external.
 Changes are saved to non-volatile memory only when one of the save options above has been selected. Any powerdown before this results in lost configuration changes.
 Selecting 'Cancel' discards unsaved changes and returns the recorder to the Operating level.
 New internal data files for enabled recording channels are created if:
– A recording channel source parameter is changed
– The primary and/or secondary sample rates and/or their sources for either process group are changed.
– The input filter type parameter for any channel is changed
– the engineering range parameter for any channel is changed
– A channel tag parameter is changed
– The number of process groups is changed
– A previously disabled channel is enabled
A warning is displayed if a configuration change will result in the creation of new data files. Select 'Yes' to accept the changes or 'No' to reject them.
 Selecting 'Save as Current Configuration' suspends recording for a short time while the new configuration is implemented.
 c Select 'Save configuration' to save any changes but to continue to use the previous configuration. d Save the configuration in internal or external storage.
e Select 'Cancel' to discard all changes and return to the Operating level.
    Note.
 A configuration file is saved with the filename '<time><date><>instrument tag>.cfg'
 When a current configuration is saved to external storage, it is also saved automatically to internal storage.
  72 IM/SM500F Rev. AA








































































   72   73   74   75   76