Page 37 - Quell-user_manual_update_321
P. 37
< Contents | APPENDIX I: Electromagnetic Compatibility Declaration
APPENDIX I
Electromagnetic Compatibility Declaration
Quell is intended for use in the electromagnetic environment specified below. The user
should ensure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions; CISPR 11 Group 1 Quell uses RF energy only for its internal function. Its
RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions; CISPR 11 Class B Quell is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonics Current Class A
Emissions IEC 61000-3-2
Fluctuations and Flicker Complies
IEC 61000-3-3
Electrostatic Discharge +/- 15kV air; +/- 8kV contact Floors should be wood, concrete or ceramic tile.
Immunity (ESD); If floors are covered with synthetic material, the
IEC 61000-4-2 humidity should be at least 30%
Radiated RF; 10 V/m; 80 MHz to 2.7 GHz 10 V/m compliance level
IEC 61000-4-3
Electric Fast Transient ±2kV power leads Mains power quality should be that of a typical
Burst Immunity domestic establishment.
IEC 61000-4-4
Fast Surge Immunity ±2kV Common mode
IEC 61000-4-5
Radio Frequency 3 VRMS (80% AM at 1 kHz);
Common Mode Immunity 150 kHz to 80 MHz
IEC 61000-4-6
Power Frequency 30 A/m 50 Hz
Magnetic Field Immunity
IEC 61000-4-8
Voltage Dip and Interrupt Complies Mains power quality should be that of a typical
Immunity IEC 61000-4-11 domestic establishment.
Quell User Manual Quell Customer Care 1-800-204-6577 or +1-781-890-9989
®