Two Yellow Collecting Waves                           English










              Description: The Communicator started but was
                           not able to complete the interrogation
                           within the time allowed.

              Action:      •  Ensure the Communicator is
                             optimally placed as described
                             in “Where to Place Your
                             Communicator” on page 13.
                           •  Face the Communicator. Sit directly
                             in front of the Communicator. Make
                             sure you are within 3 feet (1 meter)
                             of the Communicator.
                           •  Remain still until the interrogation is
                             complete. Do not move away from
                             the Communicator.
                           •  Move any wireless electronic
                             products (such as cordless or
                             cellular phones or baby monitors)
                             that are within 3 feet (1 meter) of the
                             Communicator.

                           To verify that troubleshooting was a
                           success:
                           •  Press the Heart button to start
                             another interrogation. If the
                             interrogation was a success, all
                             three Collecting Waves will light
                             green for 2 minutes.


                                     ®