PREVENTATIVE MAINTENANCE SERVICE REPORT


        Facility       Florida Hospital Orlando                 Date            12-15-17

        Address        Address                                  Technician      Anthony Voight
        City/State     Orlando, FL                              PO#
        Facility Contact/Supervisor      Darrell Huckaby                     Department       Engineering/Maint.

        Phone#   407-123-4567            Fax#     407-123-4567               Email
           Equipment Description        Equipment  Manufacturer      Model #/ Part # /Ser #     Equipment Location
        Lakeside
        Med Air System Triplex Recip.                                LPS-20T-D200-TCY
        Base Mount Vertical Tank 20hp   Beacon Hitachi (12/9/2002)   59021                     Plant Ops
        System Voltage   460/3/60   Wiring Dia. Present   Y   System Operating Pressure  Lead   96   #2   80   #3   75   #4
        ☐   Inspected/Changed Oil & Filter                                                                                                  ☒   Tested Alternation & Minimum Run Time
        ☐   Inspected/Changed Exhaust filters/Drained Drip legs                                                                                           ☒   Tested All Local Alarms
        ☒   Inspected Inlet Filters                                                                                ☒   Inspected Control Cabinet/Tightened Connections
        ☒   Inspected Dryer Pre-Filters                                                                                         ☒    Inspected/Tested Dew Point  -11.0F  CO Monitor 0ppm
        ☒   Inspected Dryer After Filters                                                                                       ☒    Room Ventilation/Ambient Temperature  75F
        ☒   Inspected Drain Floats                                     ☒ Pump1 Amps   L1    22.6            L2   19.2   L3   22.1   HR   15711   Temp   130F

        ☒   Inspected Dryer Desiccant                                                       ☒ Pump2 Amps     L1   20.5   L2   22.4   L3   22.2   HR   15867   Temp   121F
        ☒   Inspected  Belts                                           ☒ Pump3 Amps   L1   21.7   L2   21.3   L3   19.6   HR   15757   Temp   128F

        ☒   Inspected all Pressure Gauges/Switches                     ☐ Pump4 Amps   L1          L2          L3          HR         Temp
        ☒   Inspected/Tested After Cooler Drains                                                                         ☒   Inspected/Adjusted Final Line Pressure Regulators
        ☒   Inspected/Tested Receiver Tank Drains                                                                     ☒   Inspected System for Unusual Noise and Vibration
        ☒   Inspected/Tested all Discharge/Inlet Check Valves                                                    ☒   Inspected System for Air Leaks
        ☒   Inspected  Intake/Exhaust  Air Locations (Roof)          ☐   Presence of Hydrocarbons/Air Tested
        Hours:                      Per Diem:                 Mileage:                    Rate:
        Comments & Recommendations