"Oil-Less” Medical Air                                                              Medical Systems


               3.0 System Operation (continued)


               3.6 Emergency Shutdown / Alarms (continued)

               Low Dryer Outlet Pressure Alarm (Optional) - For units with this feature, there is an adjustable pressure
               switch located upstream of the dryer check valve that will alarm if the pressure in that line drops below the 75
               psig factory setpoint.  In the event of an alarm, the “dryer fault” light will come on and a remote set of contacts
               will change condition (refer to the wiring diagram that came with the unit for terminal numbers).  Pushing the
               illuminated pushbutton can reset the dryer fault alarm.