4.8 Geometric Quick-Connect Key Style Frontbody Assembly
Chapter 4
Figure 4-9. Geometric Quick-connect Key Style Frontbody Assembly
NOTE:
For Waste Anesthetic Gas Disposal (WAGD), and AGSS service only, the barrel
assembly is permanently attached to the frontbody keying disc assembly.
4. Remove all of the internal parts from the twist-lock replaceable barrel assembly
(B) by performing the following:
a. Carefully remove the frontbody piston spring (C).
b. Remove the frontbody piston (D).
c. Remove the frontbody piston O-ring (E) from the frontbody piston (D).
d. Remove the O-ring (F) and washer (G).
Replacement
1. Inspect the parts of the frontbody assembly for excessive wear or damage.
2. Replace defective parts as necessary.
3. To install the replacement Geometric Quick-Connect key style frontbody
assembly, reverse the removal procedure.
4. To ensure proper operation of the Geometric Quick-Connect key style frontbody
assembly, perform the “Function Checks” on page 2–3.
Page 4-18 DiamondCare® Recessed Wall Outlet and Electrical Accessories (MAN01-045)