Cleaning
                Chapter 6


                Cleaning




                              WARNING:
                       Follow the product manufacturer’s instructions. Failure to do so could result
                       in personal injury or equipment damage.




                              CAUTION:
                       Do not use harsh cleaners/detergents such as scouring pads and heavy duty
                       grease removers, or solvents such as toluene, xylene, and acetone.
                       Equipment damage could occur.




                              CAUTION:
                       Do not use oil or grease on or around the outlet. Doing so could result in
                       equipment damage. Use only lubricants approved for oxygen service, such
                       as Krytox® 6PL205.

                If there is no visible soilage with possible blood or body fluids, we recommend that you
                clean the unit with a mild detergent and warm water. If disinfection is desired, you may use
                a combination cleanser/disinfectant as explained in “Disinfection” below.

                Steam Cleaning

                Do not use any steam cleaning device on the DiamondCare® Recessed Wall Outlets.
                Excessive moisture can damage mechanisms in this unit.

                Cleaning Hard to Clean Spots

                To remove difficult spots or stains, we recommend that you use standard household
                cleansers and a soft-bristled brush. To loosen heavy, dried-on soil, you may first need to
                saturate the spot.                                                                              6

                Disinfection

                When there is visible soilage, we recommend that you disinfect the unit with a
                tuberculocidal disinfectant. (For customers in the United States, the disinfectant should be
                registered with the Environmental Protection Agency.)

                Dilute disinfectants as specified on the manufacturer’s label.












                DiamondCare® Recessed Wall Outlet and Electrical Accessories (MAN04-045)  Page 6-2