"Oil-Less” Medical Air Medical Systems
Appendix C - Ceramic Dew Point Monitor
C.5 Troubleshooting
WARNING:
Before removing the dew point sensor, verify that line pressure has been valved off or reduced to
atmospheric pressure.
Before servicing the dew point sensor, do the following:
1. Unplug sensor(s) or turn off monitor.
2. Depressurize the dewpoint sensor.
Note: Remote alarms will be activated.
Problem Possible Causes Solution
Slow system response Insufficient flow through dew Check flow
point sensor.
No power No incoming power. Verify line power is being
supplied.
Erratic display Unit defect. Contact BeaconMedæs.
LED display does not work or Display board defect Contact BeaconMedæs.
has missing segments
Loose connections Check connections
High Dew point Air is not being dried. Verify that dryer is not being
bypassed.
Faulty monitor or sensor. Replace monitor or sensor
C-5