DIP SWITCH SETTINGS FOR LEFT & RIGHT BANK
Field Adjustments
Left Bank Select
HP Mode Liquid Mode
ON ON
1 2 3 4 5 6 7 8 1 2 3 4 5 6 7 8
S3 S3
High Pressure mode for Amico Model #: M2-X-MAN-07B (2500 Psi)
Set Dip switch # 3 to the ON position.
This will enable gas cylinders to be attached to the left bank of the manifold.
Liquid Mode for Amico Model #: M2-X-MAN-07C (500 Psi)
Set Dip switch # 3 to the OFF position.
This will enable liquid cylinders to be attached to the left bank of the manifold
Right Bank Select
HP Mode Liquid Mode
ON ON
1 2 3 4 5 6 7 8 1 2 3 4 5 6 7 8
S3 S3
High Pressure mode for Amico Model #: M2-X-MAN-07B (2500 Psi)
Set Dip switch # 5 to the ON position.
This will enable gas cylinders to be attached to the right bank of the manifold.
Liquid Mode for Amico Model #: M2-X-MAN-07C (500 Psi)
Set Dip switch # 5 to the OFF position.
This will enable liquid cylinders to be attached to the right bank of the manifold
Example 1 Example 2
HP x Liquid Mode Liquid x HP Mode
ON ON
1 2 3 4 5 6 7 8 1 2 3 4 5 6 7 8
S3 S3
Dip Switches # 4, 6 & 7 are not used. Leave in the ON position.
Dip Switch # 8 should remain in the ON position and should never be tampered with. PAGE 19