        DIP SWITCH SETTINGS FOR  LEFT &  RIGHT BANK
 Field Adjustments

                                                   Left Bank Select


                                          HP Mode                      Liquid Mode
                                    ON                            ON




                                      1  2  3  4  5  6  7  8         1  2  3  4  5  6  7  8
                                     S3                             S3

        High Pressure mode for Amico Model #:  M2-X-MAN-07B (2500 Psi)
        Set Dip switch # 3 to the ON position.
        This will enable gas cylinders to be attached to the left bank of the manifold.

        Liquid Mode for Amico Model #:  M2-X-MAN-07C (500 Psi)
        Set Dip switch # 3 to the OFF position.
        This will enable liquid cylinders to be attached to the left bank of the manifold


                                                 Right Bank Select


                                           HP Mode                     Liquid Mode
                                    ON                            ON




                                       1  2  3  4  5  6  7  8        1  2  3  4  5  6  7  8
                                      S3                            S3

        High Pressure mode for Amico Model #:  M2-X-MAN-07B (2500 Psi)
        Set Dip switch # 5 to the ON position.
        This will enable gas cylinders to be attached to the right bank of the manifold.


        Liquid Mode for Amico Model #:  M2-X-MAN-07C (500 Psi)
        Set Dip switch # 5 to the OFF position.
        This will enable liquid cylinders to be attached to the right bank of the manifold


                                        Example 1                      Example 2
                                    HP x Liquid Mode               Liquid x HP Mode
                                  ON                             ON




                                     1  2  3  4  5  6  7  8         1  2  3  4  5  6  7  8
                                    S3                            S3

        Dip Switches # 4, 6 & 7 are not used. Leave in the ON position.
        Dip Switch # 8 should remain in the ON position and should never be tampered with.                    PAGE   19