Page 80 - Med Plaza and Cancer Center
P. 80
ZONE VALVE/ AREA ALARM / OUTLET PERFORMANCE/GAS CONCENTRATION AND PURITY INSPECTION DOCUMENT
Facility Florida Hospital Orlando-Med Plaza Inspector Anthony Voight
City/State Orlando, FL Date 12-4-17 Page 1 of 2
Contractor n/a Facility Contact Darrell Huckaby
Area Served OR 1 Floor/Bldg 4 Floor Med Plaza
th
Zone Valve Location for this Area Outside OR1 ZVB Manufacturer Medaes
FL Hosp. Tag # MA/VA/O2/NO/C2/N2-P-5PC-04-B-Z-W-01
Location IAW NFPA 99 Pass Labeled Pass Gauges Pass Leak Test Pass
Test Gauge O2 52 AIR 49 VAC 25 N2O 53 N2 167 CO2 53
Pressure vs.
ZVB Gauge 50 48 25 53 175 53
Valve Size 1/2 1/2 1 1/2 1/2 1/2
Comments:
Gas/Pipeline Purity Oxygen Med Air N2O CO2 Comments
Hydrocarbons <1 PPM <1 PPM <1 PPM <1 PPM
Dew Point -46 23 -27 -17
Carbon Monoxide <1 PPM <1 PPM <1PPM <1 PPM
Carbon Dioxide <1 PPM 370 PPM <1 PPM 5000PPM
Gas Concentration 99.9%O2 21.0 (%O2) 99.9%N2O 99.9%CO2
Particulates <1 <1 <1 <1
Odor/Discoloration None None None None
Test Location OR1 OR1 OR1 OR1
Area Alarm Panel Location for this Area In OR nurses station
Location IAW NFPA 99 Pass Labeled Pass Gauges Pass Leak Test Pass
Area Alarm Panel Manufacture Data Medaes Analog (GP-L-5PC-04-5G-SA4) (Vac GP-L-5PC-04-2G-ZO1)
GAUGE LOW HIGH TEST
GAS/VAC PSI ALARM ALARM VISUAL AUDIBLE SWITCH SENSOR COMMENTS PASS/FAIL
OXY 51 40 60 P P P SS PASS
AIR 51 40 60 P P P SS PASS
VAC 23 12 - P P P SS PASS
N2O 54 40 60 P P P SS PASS
N2 172 130 200 P P P SS PASS
CO2 51 46 60 P P P SS PASS
PS= Patient Side SS= Source Side NW= Not Working CNT= Could Not Test NA = Not Applicable NP = Not Present
Corporate Office ∙ 4518 Clements Rd. Lakeland Florida 33811 Phone: 863-646-7555 ∙ Fax 863-646-2999
“Medical Gas Pipeline Equipment Service, Sales & Verifications”
©Daniel Voight 2009