"Oil-Less” Scroll Medical Air                                                      Medical Systems


               2.0 Installation



               2.1 Inspection Upon Receiving

                                         ®
               The condition of the LifeLine  Scroll Medical Air system should be carefully inspected upon delivery. Any
               indication of damage by the carrier should be noted on the delivery receipt, especially if the system will not be
                                                       ®
               immediately uncrated and installed. LifeLine  Scroll systems may remain in their shipping containers until
                                            ®
               ready for installation.  If LifeLine  Scroll systems are to be stored prior to installation, they must be protected
               from the elements to prevent rust and deterioration.

               DO NOT REMOVE the protective covers from the inlet and discharge connection ports of the unit until they
               are ready for connecting to the hospital’s pipeline distribution system.

               2.2 Handling


                                                        !!WARNING!!
                     USE APPROPRIATE LOAD RATED LIFTING EQUIPMENT AND OBSERVE SAFE LIFTING
                                            PROCEDURES DURING ALL MOVES.


               The compressor package can be moved with either a forklift or dollies. Be sure that the orange spacers used to
               prevent the compressor and  motor  assemblies  from  floating are in place. These spacers will prevent
               unnecessary movement of the compressor and motor while moving and mounting the unit. Keep all packing in
               place during installation to minimize damage. Walk along the route the unit must travel and note dimensions
                                                  ®
               of doorways and low ceilings. LifeLine  Scroll 2-7½ hp systems are designed to go through 36” doorways.
                       ®
               LifeLine   Scroll  10-15  hp  systems must be separated to go through 36" doorways.  If this is necessary,
               carefully  label  all electrical connections that are removed for easier re-assembling at the final destination.
               Units should be placed to ensure easy access to perform maintenance and high visibility of indicators and
               gauges.

               2.3 Location

                           ®
               The LifeLine  Scroll Medical Air system should be installed indoors in a clean, well-ventilated environment.
               Areas of excessive dust, dirt or other air-borne particulate should be avoided.

               Secure the package to a flat, level surface capable of supporting the weight and forces of the unit. Make sure
               that the main base is not bowed, twisted, or uneven. Because of the internal flexible hose connections and
               spring isolators, no special foundation is required. However, the unit base must be securely bolted using all
               mounting holes provided in the base. If a raised concrete pad is used, the base must not overhang the concrete
               pad. A method to drain away moisture is necessary. If a gravity drain is not available, a connection to a drain
               is necessary. After securing the unit to the floor, remove the orange spacers from under the compressor/motor
               structure.