"Oil-Less” Scroll Medical Air                                                       Medical Systems


               Appendix C - Ceramic Dew Point Monitor



               C.2 Specifications

               C.2.1 Monitor

               System Power Requirements: 10 Watts, 110 VAC (90-126 VAC), 50/60 Hz

               System Operational Ambient Conditions:  -20° to 50°C continuous, 0 to 90% RH

               Electronic Accuracy (Repeatability, Hysteresis, Linearity):  Dew point ± 0.5°C (± 1°F)

               Resolution of Display:  Dew point: 0.1°; CO: 1 ppm

               Dew point Measurement Range:  -80° to 20°C (-112° to 68°F)

               C.2.2 Sensor

               Sensor Accuracy (Repeatability, Hysteresis, Linearity):  ± 2°C (± 4°F) (For a 2-year period, with clean
               hydrocarbon-free air supply.  BeaconMedæs recommends quarterly verification of air quality.)

               Operating Pressure Range: 0 to 150 psig (0 to 1,034 kPa)

               Air Consumption: 0.2 - 0.5 LPM (0.42 - 1.0 SCFH) with 50 psi line pressure