ZONE VALVE/ AREA ALARM / OUTLET PERFORMANCE/GAS CONCENTRATION AND PURITY INSPECTION DOCUMENT

          Facility       Florida Hosp. Orlando – Women’s Center   Inspector    Joseph Voight
         City/State      Orlando, FL                              Date         11-28-17   Page      2 of 4
                                                                                  th
         Area Served     Patient Rooms 10929-10936                Floor/Bldg   10  Floor
        O2, MEDICAL AIR, N2O, CO2 Minimum Flow Requirement with 5 psig pressure drop = (3.5 SCFM/100 LPM Minimum)
        O2 & MEDICAL AIR in Critical Care Areas shall have a Transient Flow of 6 SCFM/170 LPM for 3 Seconds
        NITROGEN OUTLETS Minimum Flow Requirement with 5 psig pressure drop = (5 SCFM/142 LPM Minimum)
        VACUUM AND EVAC/WAGD OUTLETS Minimum Flow Requirement without pressure drop below 12” on adjacent inlet = (3 SCFM/85 LPM)
        All Outlets/Inlets are listed from Left to Right, Top to Bottom.
                              st
        Outlet Type/Series Legend- 1  Digit=Manufacturer; A=Amico, B=Beacon/Medaes/Ohmeda/Ohio, C=Chemetron/Allied/NCG/Oxequip,
        O=Ohio/Squire, G= Gentec, H=Hill-Rom, T= Tri-tech
         nd
        2  Digit=Type; W=Wall, C=Console, G= Gas Block, H= Hose Drop/Ceiling
         rd
        3  Digit=Connection; O=Ohmeda, C=Chemetron, D=DISS, P=Puritan Bennett, X=Oxequip
         th
        4  Digit=Series; B=Series B, 1= Diamond 1, 3= Diamond 3/Gemini, 4=Series 400, 5=Series 500, H= Horizon Gas Block, I=Integris Gas Block
        D=Diamondcare, S= Schrader, P= Geometric Shape, O= Oxequip
         th
        5 Digit=Manufactured Assembly; B= Boom/Pendant, C=Ceiling Colum, F= Floor Pedestal, H= Headwall, I=ICU Column, P=Patient Console
        Deficiencies Legend-
        LWAI = Leaks with Adapter In        LWOAI = Leaks without an Adapter In         ADNL = Adapter Does Not Latch
        LF = Low Flow                       ADND = Adapter Does Not Detach             NLP = Not Labeled Properly
        VAI=Visual and Audible Inspection    CNT = Cannot Test (Due to Infection Control/no access)
                                 Flow        Adj Vac     Transient Flow        Outlet
          Room      Gas/Vac    SCFM/LPM     Pressure   6 SCFM O2 & AIR  Type/Series       Comments/Deficiencies
         10929     Vac                                                    ACCB          CNT 10929
                   O2                                                     ACCB
                   O2                                                     ACCB
                   O2                                                     ACCB
                   Vac                                                    ACCB
                   Vac                                                    ACCB
                   Air                                                    ACCB
                   Air                                                    ACCB
                   Air                                                    ACCB
         10930     Vac                                                    ACCB          CNT 10930
                   O2                                                     ACCB
                   O2                                                     ACCB
                   O2                                                     ACCB
                   Vac                                                    ACCB
                   Vac                                                    ACCB
                   Air                                                    ACCB
                   Air                                                    ACCB
                   Air                                                    ACCB
         10931     Vac        >3           >12”                           ACCB
                   O2         >3.5                     >6                 ACCB
                   O2         >3.5                     >6                 ACCB
                   O2         >3.5                     >6                 ACCB
                   Vac        >3           >12”                           ACCB
                   Vac        >3           >12”                           ACCB
                   Air        >3.5                     >6                 ACCB
                   Air        >3.5                     >6                 ACCB
                   Air        >3.5                     >6                 ACCB





        NOTES/DISCREPANCIES:





              Corporate Office ∙ 4518 Clements Rd. Lakeland Florida 33811 Phone: 863-646-7555 ∙ Fax 863-646-2999
                                 “Medical Gas Pipeline Equipment Service, Sales & Verifications”
                                                      ©Daniel Voight 2009