ZONE VALVE/ AREA ALARM / OUTLET PERFORMANCE/GAS CONCENTRATION AND PURITY INSPECTION DOCUMENT
Facility Florida Hosp. Orlando – Women’s Center Inspector Joseph Voight
City/State Orlando, FL Date 11-28-17 Page 1 of 5
Contractor n/a Facility Contact Darrell Huckaby
th
Area Served Patient Rooms 12917-12928 Floor/Bldg 12 Floor
Zone Valve Location for this Area Corridor wall across from Rm 12919 ZVB Manufacturer Amico – 4 Valves
FL Hosp. Tag# O2 000000244435 AIR 00000024434 VAC 000000244433
Location IAW NFPA 99 Pass Labeled Pass Gauges Pass Leak Test Pass
Test Gauge O2 52 AIR 49 VAC 22
Pressure vs.
ZVB Gauge 55 53 22
Valve Size ¾” ¾” 1 ½”
Comments:
Gas/Pipeline Purity Oxygen Med Air Comments
Hydrocarbons <1 PPM <1 PPM
Dew Point -44F 5F
Carbon Monoxide <1 PPM <1 PPM
Carbon Dioxide 0 PPM 120 PPM
Gas Concentration 99%O2 21.8(%O2)
Particulates None None
Odor/Discoloration None None
Test Location 12920 12920
Area Alarm Panel Location for this Area Nurses station wall across from Rm 12922
Location IAW NFPA 99 Pass Labeled Pass Gauges Pass Leak Test Pass
Area Alarm Panel Manufacture Data Amico LCD ADV# 000000244427
GAUGE LOW HIGH TEST
GAS/VAC PSI ALARM ALARM VISUAL AUDIBLE SWITCH SENSOR COMMENTS PASS/FAIL
OXY 52 40 60 P P P PS PASS
AIR 49 40 60 P P P PS PASS
VAC 19 12 NA P P P PS PASS
PS= Patient Side SS= Source Side NW= Not Working CNT= Could Not Test NA = Not Applicable NP = Not Present
Corporate Office ∙ 4518 Clements Rd. Lakeland Florida 33811 Phone: 863-646-7555 ∙ Fax 863-646-2999
“Medical Gas Pipeline Equipment Service, Sales & Verifications”