ZONE VALVE/ AREA ALARM / OUTLET PERFORMANCE/GAS CONCENTRATION AND PURITY INSPECTION DOCUMENT
Facility Florida Hosp. Orlando – Women’s Center Inspector Anthony Voight
City/State Orlando, FL Date 11-29-17 Page 1 of 5
Contractor n/a Facility Contact Darrell Huckaby
th
Area Served NICU 4905-4912, 4931-4934 Floor/Bldg 4 Floor
Zone Valve Location for this Area Across from Room 4912 ZVB Manufacturer Beacon
FL Hosp. Tag # O2-000000311176 AIR-000000311177 VAC-000000311178
Location IAW NFPA 99 Pass Labeled Pass Gauges Pass Leak Test Pass
Test Gauge O2 52 AIR 48 VAC 20
Pressure vs.
ZVB Gauge 53 50 20
Valve Size ¾” ¾” 1 ¼”
Comments:
Gas/Pipeline Purity Oxygen Med Air Comments
Hydrocarbons <1 PPM <1 PPM
Dew Point -44F 5F
Carbon Monoxide <1 PPM <1 PPM
Carbon Dioxide 0 PPM 120 PPM
Gas Concentration 99%O2 21.8(%O2)
Particulates None None
Odor/Discoloration None None
Test Location 4907 4907
Area Alarm Panel Location for this Area Across from Room 4912
Location IAW NFPA 99 Pass Labeled Pass Gauges Pass Leak Test Pass
Area Alarm Panel Manufacture Data Beacon Total Alert Infinity ADV# 000000311174
GAUGE LOW HIGH TEST
GAS/VAC PSI ALARM ALARM VISUAL AUDIBLE SWITCH SENSOR COMMENTS PASS/FAIL
OXY 53 40 60 P P P SS PASS
AIR 50 40 60 P P P SS PASS
VAC 22 12 NA P P P SS PASS
PS= Patient Side SS= Source Side NW= Not Working CNT= Could Not Test NA = Not Applicable NP = Not Present