ZONE VALVE/ AREA ALARM / OUTLET PERFORMANCE/GAS CONCENTRATION AND PURITY INSPECTION DOCUMENT

           Facility    Florida Hosp. Orlando – Women’s Center  Inspector  Anthony Voight
           City/State   Orlando, FL                           Date       11-29-17        Page       1 of 5

           Contractor  n/a                                    Facility Contact   Darrell Huckaby
                                                                           th
          Area Served   NICU 4905-4912, 4931-4934             Floor/Bldg   4  Floor

           Zone Valve Location for this Area   Across from Room 4912       ZVB Manufacturer     Beacon

          FL Hosp. Tag #       O2-000000311176   AIR-000000311177   VAC-000000311178
          Location IAW NFPA 99   Pass          Labeled    Pass      Gauges   Pass       Leak Test   Pass
           Test Gauge   O2    52   AIR    48   VAC    20
           Pressure vs.
           ZVB Gauge          53          50          20

          Valve Size       ¾”          ¾”         1 ¼”
          Comments:


         Gas/Pipeline Purity  Oxygen       Med Air      Comments
         Hydrocarbons              <1 PPM        <1 PPM
         Dew Point                    -44F           5F
         Carbon Monoxide           <1 PPM        <1 PPM
         Carbon Dioxide             0 PPM       120 PPM
         Gas Concentration         99%O2      21.8(%O2)
         Particulates                None          None

         Odor/Discoloration          None          None
         Test Location               4907          4907

         Area Alarm Panel Location for this Area         Across from Room 4912

         Location IAW NFPA 99       Pass                 Labeled    Pass   Gauges     Pass     Leak Test   Pass
         Area Alarm Panel Manufacture Data         Beacon Total Alert Infinity ADV# 000000311174
                     GAUGE     LOW      HIGH                        TEST
          GAS/VAC      PSI    ALARM    ALARM    VISUAL   AUDIBLE   SWITCH   SENSOR        COMMENTS        PASS/FAIL
            OXY        53       40       60       P        P         P         SS                           PASS
             AIR       50       40       60       P        P         P         SS                           PASS
            VAC        22       12       NA       P        P         P         SS                           PASS
           PS= Patient Side   SS= Source Side   NW= Not Working   CNT= Could Not Test   NA = Not Applicable    NP = Not Present