Introduction
                                                                                                 Chapter 1
                All outlets are color-coded and labeled according to NFPA and CSA standards for safety. All
                outlets are listed by UL and CSA.

                DiamondCare® outlets for all pressurized medical gases are rated for a maximum service          1
                pressure of 200 psig. They contain both a secondary check valve and a primary check valve.
                DiamondCare® vacuum inlets consist of a primary check valve only. The primary check
                valve (located in the frontbody) prevents gas flow when an outlet adapter or patient
                treatment device is removed from the outlet. The secondary check valve (located in the
                rough-in assembly) prevents gas flow when the frontbody is removed from a pressure gas
                rough-in assembly.

                Operating Precautions

                Wall Outlets perform in conformity with the description contained in this service manual
                and accompanying labels and/or inserts when installed, assembled, operated, and repaired in
                accordance with the instructions provided. Outlets must be checked periodically. Refer to
                CGA Pamphlet E-10 (1999 Edition), Maintenance of Medical Gas and Vacuum Systems in
                Health Care Facilities. If an outlet does not work, the outlet should not be used. Broken,
                missing, plainly worn, distorted, or contaminated parts should be replaced immediately. If
                repair or replacement become necessary, BeaconMedæs recommends that a phone call or
                written request for service advice be made to BeaconMedæs Customer Service. This
                product and any of its components should be repaired in accordance with written
                instructions provided by BeaconMedæs.




                              WARNING:
                       Components of outlets for vacuum service must never be interchanged or
                       used in outlets delivering pressurized gas service. Possibly injury or
                       equipment damage could occur.

                Components of outlets for vacuum service must never be interchanged or used in outlets
                delivering pressurized gas service. The outlet should not be altered without the prior written
                approval of BeaconMedæs. The user of this product shall have the sole responsibility for
                any malfunction that results from improper use, faulty installation, improper repair, damage,
                or alteration.