ZONE VALVE/ AREA ALARM / OUTLET PERFORMANCE/GAS CONCENTRATION AND PURITY INSPECTION DOCUMENT

           Facility    Florida Hosp. Orlando – Women’s Center  Inspector  Anthony Voight
           City/State   Orlando, FL                           Date       11-29-17        Page       1 of 2

           Contractor                                         Facility Contact   Darrell Huckaby
                                                                           th
          Area Served   RT Equipment Room 5168D               Floor/Bldg   5  Floor

           Zone Valve Location for this Area   Next to Room 5168D          ZVB Manufacturer     Beacon

          FL Hosp. Tag #       OXY-000000311143   AIR-000000311142
          Location IAW NFPA 99   Pass          Labeled    Pass      Gauges   Pass       Leak Test   Pass
           Test Gauge   O2    52   AIR    48
           Pressure vs.
           ZVB Gauge          53          46

          Valve Size       ½”          ½”
          Comments:


         Gas/Pipeline Purity  Oxygen       Med Air      Comments
         Hydrocarbons              <1 PPM        <1 PPM
         Dew Point                    -44F           5F
         Carbon Monoxide           <1 PPM        <1 PPM
         Carbon Dioxide             0 PPM       120 PPM
         Gas Concentration         99%O2      21.8(%O2)
         Particulates                None          None

         Odor/Discoloration          None          None
         Test Location                 RT            RT

         Area Alarm Panel Location for this Area         NO ALARM FOR THIS AREA