Page 148 - Live-cellanalysis handbook
P. 148
Live-Cell Analysis Handbook — Third Edition
The following steps should be performed in a biological safety cabinet:
STEP 1 STEP 2
• Remove top of WoundMaker. • Insert plate (containing cells & media) into base plate holder.
• Set top in empty wash boat. • Remove plate cover.
STEP 3 STEP 4
• Replace pin block by guiding the rear dowels of pin block • Push and hold the black lever.
into the rear holes of the base plate.
• Do not push down.
STEP 5 STEP 6
• Lift pin block while continuing to hold the black lever down. • Replace plate cover.
• Wash wells (up to two washes).
• Add treatment conditions.
• Put into the IncuCyte and start acquiring data and images.