Page 20 - Differential Diagnosis in Small Animal Cytology, The Skin and Subcutis
P. 20

Choice and Correct Use of the Microscope

                                                                                             7
             For the routine clinical microscope, an Abbe 0.9 is considered adequate and is normally present
             on most standard microscopes. More advanced microscopes with fluorite objectives usually
  VetBooks.ir  carry a higher-corrected condenser.

             Microscope illumination
             For a very long time, microscopes were equipped with tungsten–halogen filament lamps. The
             only disadvantage of these lamps was the short lifespan caused by the high temperatures reached
             when in use. Over the past few years, microscopes have moved to LED illumination, giving a
             bright white light without heat dissipation. However, the colour characteristics of LED lights are
             not directly comparable to halogen lights, especially with earlier LED  illumination. For this rea-
             son, an adaptation time is recommended when changing from halogen to LED lighting, as the
             appearance of some stains may look different depending on the light source.

             1.3  Setting up a Microscope

             There is only one way a microscope should be set up for any given objective. Deviating from this will only
             reduced the quality of the image. When setting up a microscope it is important to carry out the following:

             •	  Position the condenser to the correct height and centration in relation to the objective.
             •	  Ensure that the diaphragm (or diaphragms) on the microscope is correctly set (see later).
             Eyepieces

             Often overlooked, the eyepieces need to be correctly set or eyestrain will occur. Most microscopes
             have one eyepiece with a focusing capability and markings showing + and – dioptre settings.


               Procedure
               •	  Focus on a specimen looking through the fixed, non-adjustable eyepiece.
               •	  Looking through the adjustable eyepiece, rotate this until the focus is reached (sharp image).



             Illumination
             •	  Koehler illumination


                Procedure
                •	  Place a slide with a good contrast specimen on the stage and use the 10× objective. Focus
                   on the specimen with the focus controls.
                •	  Move the condenser using the condenser rack and pinion controls under the stage until
                   it is about 0.5 cm from the top.
                •	  Close the field diaphragm. At this point the leaves of the diaphragm (which may not be
                   in the centre of the field) will become visible.
                •	  Move the condenser up and down until the leaves are in focus.
                •	  Use a knurled screw or Allen key (usually in the front of the condenser mount, at the left
                   or right) to centre the image of the diaphragm.
                •	  Open the field diaphragm until it is just out of the field of view.
                In this way the condenser will be at the correct height and centred in the optical pathway, allowing
                the best illumination of the specimen to achieve the best possible image quality.
   15   16   17   18   19   20   21   22   23   24   25