Page 138 - The Veterinary Laboratory and Field Manual 3rd Edition
P. 138

The selection, use, maintenance and quality control of laboratory equipment and supplies   107


                GLycEroL                                   objects to be sterilized in the same way as for
                Glycerol in a 50% solution will kill contami-  the autoclave method. Cotton-wool plugs, if
                nating microorganisms. It is used for the   used, should not be too thick, otherwise the
                preservation of certain viruses and bacteria that   hot air cannot penetrate.
                are not affected by the glycerol.        •  Moist heat: A common method of wet steriliza-
                                                           tion is achieved by using an autoclave (steam
                PHEnoLIcS                                  sterilizer) at a temperature of 121°C (15 psi)
                •  Lysol is a powerful phenolic antiseptic used   for 20 min or at 110°C (10 psi) for 40 min. A
                  mainly for disinfection of discarded culture   temperature of 100°C (boiling) will kill non-
                  plates, contaminated pipettes and other   sporing organisms within 10 min. Most spores
                  potentially infectious material. Solutions of   will be killed in 30 min at 100°C but some
                  1 to 9% are generally used.              spores can resist boiling for several hours. The
                •  Hycolin: a phenol derivative used at a 1% solu-  addition of 2% sodium carbonate increases the
                  tion for all forms of laboratory disinfection.  bactericidal effect of boiling water and spores
                                                           that resist boiling water for 10 h have been
                Always read the manufacturer’s instructions
                before  using a chemical disinfectant. Used   killed in 30 min this way. This method is suit-
                incorrectly, all disinfectants can be poten-  able for sterilizing contaminated instruments
                tially hazardous to laboratory staff.      following post-mortems so that they may be
                                                           used again quickly. It is unsuitable if instru-
                                                           ments are to be stored in a sterile condition.
                Sterilization                            •  Biological fluids may be sterilized by heating
                Laboratory ware and instruments may need to   them in a water bath at 56°C for periods of
                be sterilized but the procedure is also important   1 h daily if necessary. The principle is called
                for the preparation of culture media and some   Tyndallization but at this low temperature
                reagents. Sterilization can be achieved using a   more than three exposures may be necessary.
                variety of methods and it is a case of choosing   This method of sterilization cannot be used
                one that is appropriate for the situation.  when the fluid potentially contains resistant
                                                           spores. It is a useful method to sterilize media
                PHySIcaL MEtHodS oF StErILIzatIon          containing protein or sugars which may be
                •  Dry heat: The application of dry heat is a simple   ‘broken down’ at higher temperatures and
                  method of sterilization providing the material   when filtration is not feasible.
                  to be sterilized is not adversely affected by the
                  heat. An example is nichrome wire loops used   StEaM StErILIzatIon
                  in bacteriology. These can be heated and ster-  Steam sterilization (autoclaving) is defined as
                  ilized in a Bunsen flame in excess of 300°C.   the destruction or irreversible inactivation of all
                  For glassware/metal articles that must remain   reproducible microorganisms under exposure to
                  dry, a hot air oven is used at a temperature of   saturated steam at 121°C (15 psi).
                  160°C. This temperature must be maintained   The following points must be observed.
                  for 1 h (2 h if the objects are heavy or bulky),
                  then the glassware is allowed to cool slowly   •  Efficient steam sterilization is assured only if
                  before removing it from the oven (when the   the steam is saturated and has unrestricted
                  temperature is around 40°C). Note that rapid   access to all contaminated areas.
                  cooling will cause contraction and unsterile   •  To prevent pressure build-up, containers or
                  air being sucked into the vessel. Prepare the   vessels must always be open.







       Vet Lab.indb   107                                                                  26/03/2019   10:25
   133   134   135   136   137   138   139   140   141   142   143