Page 428 - Clinical Small Animal Internal Medicine
P. 428

396  Section 5  Critical Care Medicine

            better  matched  to  the  patient’s  intrinsic  efforts.  The   Inspiratory:Expiratory (I:E) Ratio and Respiratory
  VetBooks.ir  “trigger” or “sensitivity” setting determines what the   Rate Settings
            ventilator will acknowledge  as a  genuine  inspiratory
                                                              An operator‐set respiratory rate is available on nearly all
            effort by the patient. Appropriate trigger sensitivity is
            essential to ensure the ventilator recognizes  actual   ventilators. A respiratory rate of 15–20 breaths per min-
                                                              ute with an inspiratory time of approximately one sec-
              respiratory efforts made by the patient while not   ond reflects typical initial ventilator settings in the
              delivering breaths as a result of minor fluctuations in   authors’ practice. These settings can then be adjusted as
            circuit pressure or flow. Appropriate trigger sensitivity   needed to meet the patient’s needs. The ratio of the dura-
            increases patient comfort, reduces dyssynchrony   tion of the inspiratory phase relative to the expiratory
            between patient effort and machine delivery, and allows   phase (I:E ratio) may be preset by the clinician or may be
            the patient to increase the respiratory rate as required   a setting that is determined automatically based on rate
            by their needs. The trigger variable can be too sensitive   and inspiratory time settings. Commonly, an I:E ratio of
            such that nonrespiratory efforts such as patient han-  at least 1:2 is utilized to increase the likelihood that
            dling and cardiac contractions may initiate breaths.   patients are able to fully exhale prior to the delivery of
            This situation should be avoided as it leads to patient   the next breath. As respiratory rates are increased, the
            discomfort, excessive mean airway pressures, breath   expiratory time will be reduced in order to accommodate
            stacking, auto‐PEEP, and greater dyssynchrony. An air-  the increased number of breaths. High respiratory rates
            way pressure drop of –2 cmH 2 O or a gas flow change of   can lead to a situation known as auto‐PEEP or intrinsic
            2 L/min are reasonable trigger sensitivity settings to   PEEP if the patient is not able to fully exhale before the
            start with. These settings can then be altered as   start of the next inspiration. Such an occurrence can eas-
            required. Some ventilators (e.g., AutoTrak on the   ily be detected by inspecting the flow scalar graphic if
            Phillips Esprit line) have proprietary forms of trigger   waveform displays are available. Failure of expiratory
            sensing that combine the flow and pressure signals into   flow to return to zero before the next breath is delivered
            an integrated variable. Evidence of superiority of this   is indicative of the presence of auto‐PEEP. To avoid this
            means of sensing patient inspiratory effort is currently   problem, it is recommended to keep an I:E ratio of 1:1 or
            lacking in the veterinary literature.
                                                              higher whenever possible. If a higher respiratory rate is
                                                              required, a shorter inspiratory time will allow mainte-
            Positive End‐Expiratory Pressure Settings         nance of an acceptable I:E ratio and is often better toler-
                                                              ated by the patient.
            Positive end‐expiratory pressure (PEEP) maintains
            pressure  in the breathing  circuit above  atmospheric
            pressure during exhalation so that the patient cannot   General Guidelines for Initial Ventilator Settings
            exhale all the way down to functional residual capacity
            (FRC). This increased expiratory pressure maintains   As discussed above, the optimal ventilator settings for a
            the lung at a higher volume, with less small airway col-  given patient cannot be predicted reliably. It is likely that
            lapse, and improves oxygenating efficiency. PEEP may   animals with hypoxemic respiratory failure will require
            also help to recruit alveoli collapsed due to atelectasis   more aggressive settings than those with hypercapneic
            or increased surface tension forces. Evidence suggests   respiratory failure. One must choose some initial set-
            that it may reduce ventilator‐induced lung injury by   tings prior to connecting the patient to the ventilator.
            altering  cyclic  stresses  and  strains  on  the  alveoli  and   First the mode of ventilation must be selected; A/C,
            peripheral airways. Typically, a small amount of PEEP   SIMV or continuous spontaneous ventilation. If A/C or
            (~2 cmH 2 O) is applied in all ventilator modes to reduce   SIMV are selected then it may be necessary to next
            atelectasis  and to account  for  the increased  work of   decide between volume control (VC) or pressure control
            breathing required simply due to the physical nature of   (PC) ventilation. The more modern and advanced the
            the ventilator circuit (e.g., valves, circuit resistance). In   ventilator, the more options one is likely to have.
            patients with significant lung disease, much higher lev-  Ventilator manufacturers are creating new proprietary
            els of PEEP may be required to improve oxygenating   ventilation modes faster than clinicans can evaluate
            ability. In continuous spontaneous ventilation, CPAP is   them. Even the most renowned respiratory therapists in
            functionally equivalent to PEEP although they are   the world  admit that they cannot  keep up with every
            often generated by different means (i.e., closure of the   mode becoming available. This situation is unlikely to
            expiratory valve at supraatmospheric pressures versus   change as equipment manufacturers are often less than
            continuous inflow to maintain a given circuit pressure   forthright about how these novel modes are designed.
            during the expiratory phase).                     The settings that need to be selected will vary with the
   423   424   425   426   427   428   429   430   431   432   433