FlexiVent ™
TECHNICAL SPECIFICATIONS
Operates on standard European power of 220/240V, 50 Hz
and United States Power of 110V, 60 Hz
Automatically operates and maintains an enclosure negative
pressure of -15 Pa (nominal)
Audible and visual alarms are set at -25 Pa and -3 Pa to
notify operator in case of pressure change outside of the
safe operating range or in case of excessive exhaust filter
pressure drop
Complies with the following Regulatory Requirements:
GAMP 5
21CFR Part 11
Machinery Directive 2006/42/EC replacing 98/37/EC
IEC 60529
CE Marking available