Page 40 - March 2020 - WT Site
P. 40

evaporate. The results are the weight of the oil residue remaining   solvent. For highest accuracy, UV Fluorescence analyzers should
          after evaporation. This method has a minimum reporting limit   be calibrated to the target oil in the process. However, when used
          of 5 ppm. Because of the evaporation step light hydrocarbons   with a generic calibration, the results can indicate the removal
          are under reports by this method. This analytical method cannot   efficiency of a treatment system by comparing samples before
          be used for light hydrocarbon processes or processes that aim   and after treatment.
          to achieve low concentrations. The figure included shows how
          the EPA method does not have any response to changing
          concentrations of a process, while an InfraRed technology does.

          The procedures for preparing a water sample for analysis can be
          manipulated to best demonstrate the performance of a water
          treatment system. For instance, if a treatment system is designed
          to remove 99% of the oil and solid droplets that are 10 microns
          and larger, the sample can be prepared to differentiate particles
          larger than 10 microns. The method might first test a sample of
          water using a solvent and acid, then test another sample which
          is first passed through a 10 micro filter, and then the acid and   UV  Fluorescence  can  operate  with  different  light  wavelengths.
          solvent used to prepare the sample. When the results of the   When analysis of light hydrocarbons or low concentrations is
          second test are subtracted from the first, the results will be the   needed, a deep UV light source, with a short wavelength should
          concentration of oil that was in droplets greater than 10 microns.  be chosen. Most crude oils and light gas condensates can be
                                                                 measured below 1 ppm. However,  when a heavy oil is to be
          InfraRed analytical methods are often used for water analysis.   measured and toluene is to be used as the extraction solvent, a
          The solvent used must be considered for the extraction efficiency,   light source with a longer wavelength should be chosen. Toluene
          operating cost and solvent disposal. If the extraction solvent   can be measured by short wavelength light sources, but is optically
          is a hydrocarbon, it must be evaporated before analysis of the   clear to longer wavelengths and therefore not measured.
          remaining oil film can be performed. Similar to the EPA 1664A
          method, the IR analysis of light hydrocarbons will not have much
          resolution at low concentrations. Non-hydrocarbon solvents can
          be used without evaporation, allowing for better sensitivity to
          light hydrocarbons. For these solvents the analysis is made in the
          solvent for a quick result.















                                                                 Conclusions
          UV Fluorescence is also common for water sample analysis. All
          common solvents (hydrocarbons and nonhydrocarbons) can be   Advances in water treatment technologies are allowing discharge
          used with UV Fluorescence. For instance, when analyzing heavy,   water quality to improve significantly. With these advances, the
          high asphaltenic crude oils, toluene can be used as an extraction   water can be re-used or recycled in ways not previously available.




          42    Water Today - The Magazine  March  2020
   35   36   37   38   39   40   41   42   43   44   45